Product Description
Kit of 96 tests
Two-site ELISA method for the measurement of coagulation Factor IX, in human citrated plasma. Learn more.
Product Characteristics
Total Assay Time: 1 hr 05 min (1 Step)
Detection Threshold: ≤ 5%
Dynamic Range: 0 to 100 ng/mL
Intra-Assay CV: 3 to 8%
Inter-Assay CV: 5 to 10%
No significant interference of Heparin up to 2 IU/mL.
Reference Material – NIBSC International Standard
Product Components
Micro ELISA Plate (12 strips of 8 wells)
2x Sample Diluent
3x Factor IX Calibrator
Plasma FIX Controls (high & low)
3x Anti-h-FIX-RP IC
1x Conjugate Diluent
1x Wash Solution
1x TMB
1 x 0.45M Sulfuric Acid
Product Application
For the measurement of Factor IX antigen in research applications.
Please Note: Results obtained should be for research purposes only and not used for patient diagnosis or treatment.